MIXED CALIBRATION ON LONG BASELINE FOR LCR AND CG-5 GRAVIMETERS
Liu Shaoming 1) ;Sun Shaoan 1) ;Lu Hongyan 2) ; and Guo Shusong 3)
1)Institute of Seismology, CEA, Wuhan 430071;2)Institute of Geophsics,CEA,Beijing 100081;3)Second Monitoring and Application Center, CEA, Xi’an 710054
Abstract The mixed calibration of 8 LCR and 4 CG5 gravimeters carried out on the long baseline including 4 high precision absolute station of Wuhan, Zhengzhou, Beijing and Shenyang from May 5th to May 20th,2010 and from Jan 18th to Jan 26th,2010 is introduced. The results show that the quadratic terms exists in five gravimeters including all the 4 CG5 gravimeter, but the precision of them is relatively low. If we only calibrate the first order term, the precision of the first order term of the five gravimeters is better than the joint calibration and for the other gravimeters it is basicallly invariant.
Key words :
LCR gravimeter
CG5 gravimeter
scale function
calibration
long baseline
Received: 01 January 1900
Corresponding Authors:
Liu Shaoming
Cite this article:
Liu Shaoming ,Sun Shaoan ,Lu Hongyan et al. MIXED CALIBRATION ON LONG BASELINE FOR LCR AND CG-5 GRAVIMETERS[J]. , 2012, 32(1): 56-59.
Liu Shaoming ,Sun Shaoan ,Lu Hongyan et al. MIXED CALIBRATION ON LONG BASELINE FOR LCR AND CG-5 GRAVIMETERS[J]. jgg, 2012, 32(1): 56-59.
URL:
http://www.jgg09.com/EN/ OR http://www.jgg09.com/EN/Y2012/V32/I1/56
[1]
QIAN Wei,YUE Jianping,SHAN Lijie,HAN Chenyu. Accuracy Analysis of Solving Earth Rotation Parameters Based on VLBI Observation Data [J]. jgg, 2021, 41(12): 1245-1248.
[2]
LI Yuzhao,YAN Haowen,YNAG Weifang,WANG Shijie. Ambiguity Resolution of BDS Long Baseline Based on Fuzzy Clustering Analysis Method [J]. jgg, 2021, 41(10): 1040-1044.
[3]
. [J]. jgg, 2020, 40(S3): 73-76.
[4]
HU Zhifei,ZHANG Kanghua,TIAN Jiayong. A Review of In-Situ Calibration Method for Borehole Strainmeter Observation [J]. jgg, 2020, 40(9): 970-975.
[5]
Lü Yongqing, XIANG Ya, ZHOU Yunyao, QI Junwei, WU Huan, PENG Jing. Study on Method of Eliminating Seismometer Step Calibration Signal [J]. jgg, 2020, 40(7): 761-764.
[6]
LANG Junjian, LIANG Xinghui, TU Haibo, LIU Leijun, HU Ming, HE Jiangang, LIU Lintao. Calibration of Scale Value for CHZ-Ⅱ Air-Sea Gravimeter [J]. jgg, 2020, 40(3): 323-326.
[7]
. [J]. jgg, 2019, 39(增2): 34-36.
[8]
. [J]. jgg, 2019, 39(增2): 133-136.
[9]
XU Wei,YU Xuexiang,JIA Xue,YAN Chao,WANG Tao. Performance Analysis of Medium-Long Baseline Solution in GPS/BDS System Based on 3D Mobile Deformation Platform [J]. jgg, 2019, 39(9): 934-941.
[10]
WEN Yaxin,DAI Wujiao. A Single Epoch Ambiguity Algorithm for Medium and Long Baseline in Deformation Monitoring [J]. jgg, 2019, 39(9): 942-946.
[11]
FAN Haopeng,SUN Zhongmiao,LIU Xiaogang, ZHAI Zhenhe. Research on the Effects of Azimuth on Slant Tropospheric Delay [J]. jgg, 2019, 39(6): 634-638.
[12]
WANG Yingjian,QIU Xuefeng,ZHANG Songtang,GAO Xinbing,LI Jianguo. Calibration and Analysis of the Gaoyakou Short Gravity Calibration Baseline [J]. jgg, 2019, 39(6): 654-656.
[13]
CAO Jinguo,LI Yongming,GAO Xu’na,QI Yufei,CHE Shibao. Adjusting Principle of CG-5 Gravimeter Tilt Sensor Parameters: Analysis and Applications [J]. jgg, 2019, 39(4): 432-436.
[14]
. Method to Improve the Measurement Accuracy of Laser Tracker Based on the Length Standard Device [J]. jgg, 2019, 39(3): 325-330.
[15]
WEI Shouchun,ZHU Yiqing,ZHAO Yunfeng,HUSAN Irxat. Impact on Gravity Data Process of Scale Factor Coefficient of CG-5 Gravimeter [J]. jgg, 2019, 39(2): 210-214.